MultiNA-MCE-202 Microchip Electrophoresis System for DNA/RNA Analysis
Instrument Details
- Make : Shimadzu Corporation, Kyoto Japan
- Model : MCE-202 MultiNA
Specifications :
- Sample rack: Compatible with 96-well PCR plate† and 12/8-strip PCR tube (Shimadzu recommended product)
- Microchip: Quartz, 23 mm separation channel length, on-chip electrodes (insert up to four microchips)
- Electrophoresis: Voltage Max. rated voltage: 1.5 kV, max. current: 250 μA
- Detection Method: LED-excited fluorescence detector (470 nm excitation wavelength) * Class 1 LED product
- Separation Size Range: 25 to 500 bp (DNA-500 Kit)
100 to 1000 bp (DNA-1000 Kit)
100 to 2500 bp (DNA-2500 Kit)
100 to 12000 bp (DNA-12000 Kit)
Up to 28S rRNA (5.0 knt) (RNA Kit) - Resolution: 5 bp (25 to100 bp), 5% (100 to 500 bp), 10% (500 to 1000 bp), 20% (1000 to 12000 bp)
- Sizing Accuracy: ±5 bp (25 to 100 bp), ±5% (100 to 500 bp), ±15% (DNA-1000, DNA-2500, DNA-12000)
- Required Sample Volume: DNA analysis: Premix mode: 2 to 10 μL (after mixing with marker solution: 6 to 30 μL), On-Chip Mixing mode: 5 to 30 μL, RNA analysis: Premix mode: 3 to 15 μL (after mixing with marker solution: 6 to 30 μL), In the Premix mode, the marker solution is mixed with the sample before loading in the instrument. In the On-Chip Mixing mode, the sample and marker solution are loaded separately and mixed on the microchip under program control
Applications: Widely used for genetic research such as:
- Genotyping and fragment analysis of plant populations
- Microbiological analysis using SSR markers
- Infectious disease analysis using SSR markers
Contact Us:
- Director, ICAR-CPRI,Shimla(H.P.)-171001
- Phone: 0177-2625073
- Mobile:+91-9418046415
- Email Id: director.cpri@icar.gov.in